![]() |
As alluded to in section 3.4, the majority of the
available magnetic diagnostics on W7-AS were unusable in our analysis
due to lack of compensation for drifts in the power supply and/or
failing sampling electronics. Figure 3.19 of the raw
magnetic signals for shot 43359 illustrates some of these problems.
The moments from the poloidal field measurements [a]-[c] clearly
suffer from non-zero signal drift, which is a result of a constant
offset voltage across the integration electronics. If the drift were
reasonably linear, as in the case of [a], then this could easily be
subtracted off after sampling. However, for [b] and [c], this drift
is clearly not purely linear and thus introduces varying degrees of
uncertainty in the signal magnitude. Here, this is more serious for
[c] due to the lower signal strength. Possible explanations (aside
from malfunctioning electronics) may be variations in the (supposedly
constant) vacuum field due to drifts in the coil currents [e]-[h]
originating in the power supply. If these were measured sufficiently
accurately, they could be compensated for during post-processing using
a series of calibration shots. The jumps in [e]-[g] indicate that
the currents are roughly constant within the signal bit resolution
(24.4 A), however the magnetic diagnostics are sensitive to field
variations in this range. We remark that the overall signal quality
of the
cosine moments is otherwise quite acceptable at
half-field (i.e. for
and
near 15 kA each, where the main
magnetic field is roughly 1.25 T) but degrades sharply at full field
(for
and
near 30 kA, where the field is roughly 2.5 T) as
shown in figure 3.20.
![]() |
The diamagnetic signals [d] were unaffected by these considerations,
however, and we demonstrate the FP procedure and show the typical
accuracy achievable using only the vacuum field information and a
diamagnetic signal to measure the plasma energy content and thus the
first principal component of the foregoing section. Of the two
possibilities, we choose the diamagnetic coil located near
over that near
(both shown in
fig. 2.6), as the latter lacks hardware compensation.
However, the compensation itself could also prove problematic, since
the compensation coil in question also picks up a signal from the
plasma itself, thus subtraction of the compensation signal
systematically reduces the measured diamagnetic signal and gives
erroneously low estimates of
. Figure 3.21
demonstrates this effect by plotting the uncompensated flux signal
from the diamagnetic coil against the compensated signal for each case
in the database -- with very slight scatter due to weak
configurational dependence, the compensated signal is consistently
13% smaller than the uncompensated one. This presents no difficulty
to FP, however, since the plasma contribution to the compensation coil
signals can be simulated using DIAGNO and the over-compensation
effect can be accurately predicted. Reconstructions can thus still be
performed directly with the experimental signal.
FP reconstructions using experimental data from the diamagnetic coil
are compared to standard NEMEC equilibrium calculations in three
cases. Note that this indicates the most pessimistic level of
accuracy achievable in magnetic reconstructions, since it assumes
knowledge only of the integral of the pressure profile and not its
shape. It essentially leads to a single average profile form on all
recovered equilibria and to discrepancies in the flux surface geometry
when the profile in question differs strongly from the database
average (see below). A fully quadratic model in the FP predictors is
used: in the notation of section 3.5, we have
, the diamagnetic flux. Note that since
the vacuum predictors in the database are chosen independently of one
another and the plasma information is itself independent of the vacuum
parameters, the usual PCA step in the FP procedure is unnecessary and
is omitted.
|
|
|
We compare FP reconstructions to standard equilibrium reconstructions
using Thomson data for three cases in figs. 3.22,
3.23 and 3.24. The key to the plots
is as follows: equilibrium parameters in the top left,
profile
in the top middle, root-mean-square (RMS) difference between FP and
standard NEMEC surfaces as a function of
in the poloidal
sections at
shown in the lower 3
plots.
The first case (fig. 3.22) is shot 31114, a high
discharge with a fairly linear pressure profile in
that is
close to the average profile shape in the FP database. The agreement
of the flux surfaces is moderate near the axis but improves steadily
towards the boundary. This is to be expected, since the Shafranov
shift (
) of the flux surfaces is dependent on the local
value (and thus on the shape of the pressure profile), which
is poorly diagnosed by external magnetic measurements towards the
plasma core. Agreement with integral quantities
,
etc. is good, however, and the plasma boundary matches that in the
standard calculation.
Shot 31119 (fig. 3.23) has a lower
and thus
the deviation of the flux surfaces from the well recovered vacuum
configuration is not as severe as for 31114. The pressure profile is
linear in
and similar to the database average. Flux surfaces
agree well towards the boundary and are still reasonably close in the
vicinity of the magnetic axis. Scalar parameters agree well also.
The limitation of having no profile information is most apparent for
the high-
shot 31901 (fig. 3.24), where the
peaked pressure profile differs strongly from the database average.
This results in a higher Shafranov shift for inner flux surfaces
(
) and thus a large discrepancy between FP
flux surfaces and the standard calculations towards the centre of the
plasma. Again, boundary agreement is within a few mm and scalar
parameters are well recovered.